服务项目 | 浏览量0 | 发布时间2013-11-10 |
品牌未填 | 所在地 | 起订≥1 立方 |
供货总量未填 | 发货0天内发货 | 有效期至2025-03-09 |
The CT2 Current Probes are designed for permanent or semi-permanent in-circuit installation. Each probe consists of a current transformer and an interconnecting cable. The current transformers have a small hole through which a current carrying conductor is passed during circuit assembly.
The P6041 Probe Cable provides the connection between the CT2 Current Transformers and a BNC oscilloscope input. A 50 Ω termination is required to terminate the cable when connected to a high-impedance (1 MΩ) oscilloscope input. One probe cable can be used to monitor several current transformers that have been wired into a circuit.
Bandwidth (typical) |
1.2 kHz to 200 MHz |
|
Rise Time |
500 ps |
|
Sensitivity (into 50 Ω) |
1 mV/mA |
|
Accuracy |
±3% |
|
Magnetizing Inductance |
7 μH |
|
Leakage Inductance |
1 nH |
|
Insertion Impedance: |
||
at 10 MHz |
0.1 Ω |
|
at 100 MHz |
0.5 Ω |
|
at 1 GHz |
|
|
Max. Bare Wire Size |
#16 wire |
|
1.32 mm (0.052 in.) |
||
Max. Bare Wire Voltage: |
||
RMS |
175 VRMS CAT I |
|
Peak |
1000 V*1 |
|
DC |
175 V |
|
Max. Peak Pulse Current |
36 A |
|
Max. Continuous Current (RMS) |
2.5 A |
|
Amp-second Product |
50 × 10-6 A*Sec |
|
L/R Time Constant (droop) |
>160 μs |
|
Propagation Delay |
6.1 ns |
*1 (<3.25% Duty Factor).